Rebasing

Rebasing

The rebasing process standardises sensor output following the stabilisation of a sensor and is a vital stage before the start of a new project if there is any possibility that sensors have been destabilised. Sensors automatically rebase when first shipped by the manufacturer but after that, including if the pod is first switched on at a location other than the monitoring site, such as in an office, the process must be triggered manually. This will give the best possible pod-to-pod comparability and ‘out-of-the-box’ accuracy.

  • Sensors must stabilise in the monitoring environment for two days before rebasing starts
  • Rebasing will not be effective if carried out on a destabilised sensor
  • Rebasing must be carried out in the monitoring environment
  • Rebasing must not be interrupted by moving the pod during the process